SECTION 3.0, OPTIONAL EQUIPMENT
3.3.2.1 Resistive Temperature Device (RTD)
An optional Resistive Temperature Device (RTD) probe is available for
use as an EXTERNAL MEDIUM TEMPERATURE sensor. The heart of
the assembly is a 1000 ohm 2 wire RTD. Consult the system wiring
diagram for connection of the RTD Assembly to the TIE.
The optional Dual Analog Input Board is required to use the RTD
assembly. See Table 3-2 for the applicable Dual Analog Input Board
jumper settings for the assembly.
Table 3-2
Dual Analog Input Board Jumpers for External Temperature
JUMPER
OPTIONS
DEFAULT
FUNCTION
JU1
RTD = 1-2
CUR = 3-4
VOL = 5-6
RTD = 1-2
Selects external input to
temperature channel between
RTD, 1-5 volts, or 4-20 mA
JU4
3-4 (1000 ohms)
3-4 (1000 ohms)
Selects the 32°F (0°C) RTD
resistor for calibration
JU7
3-4 (1000 ohms)
3-4 (1000 ohms)
Selects the 1000 ohm RTD
for operations
3.4 LINK ROD
For stacks with an annulus or thick outer walls that must be bridged by the
Ultraflow 150
Purge Nozzle Assemblies, Teledyne Monitor Labs offers
the Link Rod option whenever the Purge Nozzle Assemblies would exceed
72 inches (183 cm) in length. Beyond this length, the standard assemblies
become unwieldy and are difficult to align in the center of their ports. The
Link Rod Assemblies replace the standard Purge Nozzles. The Link Rod
Assemblies consist of a transducer housing with skids to center it in the
port, and a series of short rods that connect or disconnect as the transducer
housing is inserted or removed from the port. The Link Rod option also
reduces the amount of removal clearance required to extract the
transducers for maintenance and cleaning. This mounting option is
complete with all hardware, flanges and seals.
Consult the Site Specification Data sheet of Appendix A for information on
the configuration of your individual monitor.
3-4
Summary of Contents for ULTRAFLOW 150
Page 2: ...DOCUMENT NO 1900 0100 01 REV E ...
Page 24: ...ULTRAFLOW 150 GAS FLOW AND TEMPERATURE MONITOR This page intentionally left blank 4 6 ...
Page 75: ...SECTION 7 0 INSTALLATION This page intentionally left blank 7 7 ...
Page 76: ......
Page 85: ...SECTION 8 0 SYSTEM CALIBRATION AND ADJUSTMENT This page intentionally left blank 8 9 ...
Page 86: ......
Page 89: ...APPENDIX A SITE SPECIFICATION DATA SHEETS ...
Page 90: ...ULTRAFLOW 150 GAS FLOW AND TEMPERATURE MONITOR This page intentionally left blank ...
Page 95: ...APPENDIX B MAINTENANCE CHECK SHEETS ...
Page 96: ... This page intentionally left blank ...
Page 101: ...APPENDIX C SPARE PARTS ...
Page 102: ... This page intentionally left blank ...
Page 105: ......
Page 106: ...APPENDIX D DRAWINGS ...
Page 107: ... This page intentionally left blank ...
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ......
Page 112: ......
Page 113: ......
Page 114: ......
Page 115: ......
Page 116: ......
Page 117: ......
Page 118: ......
Page 119: ......
Page 120: ......
Page 121: ......
Page 122: ......
Page 123: ......
Page 124: ......
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 132: ......
Page 133: ......
Page 134: ......
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ......
Page 139: ......
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ...APPENDIX E ENHANCED SERIAL PORT COMMUNICATION PROTOCOL ...