Scale setting
Position setting
(Divs)
Offset setting
5
Termination
setting
Generator
setting
Accuracy limits
120 mV
0
0 V
0 V
0 V
–14.5 mV to +14.5 mV
+3.4 V
+3.5 V
+3.4 V
+3.3717 V to +3.4283 V
-3.4 V
-3.5 V
-3.4 V
–3.4283 V to –3.3717 V
140 mV
0
–6 V
0 V
–6 V
–6.0525 V to +5.9475 V
0 V
0 V
0 V
–16.5 mV to +16.5 mV
+6 V
0 V
+6 V
–5.9475 V to +6.0525 V
200 mV
0
–6 V
0 V
–6 V
–6.0585 V to –5.9415 V
0 V
0 V
0 V
–22.5 V to +22.5 V
+6 V
0 V
+6 V
+5.9415 V to +6.0585 V
400 mV
0
–6 V
0 V
–6 V
–6.0785 V to –5.9215 V
0 V
0 V
0 V
–42.5 mV to +42.5 mV
+6 V
0 V
+6 V
+5.9215 mV to +6.0785 mV
d. Display the test signal:
■
Display the
Vertical Setup
menu.
■
Use the keypad to set the Termination Voltage, Position, and Offset
vertical settings as listed in the table for the current vertical scale
setting. The baseline level may move off screen.
■
Set the generator so that its Sense value and polarity matches the
generator setting in the table for the vertical scale, termination
voltage, position, and offset settings you have made.
NOTE.
It’s important to use the Sense reading of the generator instead of
the Output level when adjusting the generator to the specified voltage
setting.
5
Set as precisely as the instrument's offset resolution permits.
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
222
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series