Specifications (MSO/DPO5000/B Series)
This chapter contains specifications for the MSO5000/B and DPO5000/B series
oscilloscopes. All specifications are guaranteed unless noted as "typical." Typical
specifications are provided for your convenience but are not guaranteed.
Specifications that are marked with the
symbol are checked in Performance
Verification.
All specifications apply to all MSO5000/B and DPO5000/B models unless noted
otherwise. To meet specifications, two conditions must first be met:
■
The instrument must have been operating continuously for twenty minutes
within the specified operating temperature range.
■
You must perform the Signal Path Compensation (SPC) operation before
evaluating the instrument's performance to the specifications. (See page 4-32,
Self Test.) If the operating temperature changes by more than 10 °C (18 °F),
perform the SPC operation again.
Analog channel input and vertical specification
Number of input channels
4 analog channels, digitized simultaneously
Input coupling
DC or AC
Input resistance selection
1 MΩ or 50 Ω
250 kΩ selectable for performance verification.
Input impedance DC coupled
1 MΩ
1 MΩ ±1%
50 Ω
50 Ω ±1%
MSO5204/B,
DPO5204/B
VSWR ≤1.5:1 from DC to 2 GHz, typical
MSO5104/B,
DPO5104/B
VSWR ≤1.5:1 from DC to 1 GHz, typical
MSO5054/B,
DPO5054B
VSWR ≤1.5:1 from DC to 500 MHz, typical
MSO5034/B,
DPO5034/B
VSWR ≤1.5:1 from DC to 350 MHz, typical
Input capacitance 1MΩ DC coupled 13 pF ± 2 pF
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series