f. Check against limits:
■
CHECK that the readout for the measurement
Mean
readout on
screen is within the limits listed for the current vertical scale and
termination/offset/generator settings. Enter the value on test record.
■
Repeat substep 2.d, reversing the polarity of the termination voltage,
offset, and generator settings as is listed in the table.
■
CHECK that the
Mean
measurement readout on screen is within the
limits listed for the current vertical scale setting and termination/
offset/generator settings. Enter the value on test record.
■
Repeat substeps 2.c through 2.f until all vertical scale settings are
checked for the channel under test. (See
on page 178.)
g. Test all channels:
Repeat substeps 2.a through 2.f for all four channels.
3. Disconnect the hookup:
a. Set the generator output to 0 V
.
b.
Disconnect the equipment from the generator output and the input
connector of the channel last tested.
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
182
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series