Model
Attenuator
≥ 4 GHz models
2X
DPO7354C
None
DPO7254C
5X
DPO7104C
5X
DPO7054C
5X
■
Check that a stable trigger is obtained.
c.
Repeat step 2, substeps 2.b and 2.c for the full bandwidth selected.
Table 18: Trigger settings for ≥ 4 GHz models
Generator
amplitude
Generator frequency
Horizontal scale
A trigger
B trigger
10 MHz
10 MHz
200 ns
150 mV
6 GHz
6 GHz
200 ps
200 mV
8 GHz
NA
200 ps
500 mV
11 GHz
9 GHz
200 ps
d.
≥ 4 GHz models:
Display the test signal:
■
Remove the attenuator.
■
Set the generator frequency to 10 MHz. Set the Horizontal SCALE as
indicated in the table. (See
Table 18: Trigger settings for ≥ 4 GHz
■
Fine adjust the generator output until the
Ch 1 Amplitude
readout
indicates the amplitude listed in the table for a frequency not yet
checked. (See
Table 18: Trigger settings for ≥ 4 GHz models
page 263.)
■
Set the generator frequency to the frequency in the table that
corresponds to the amplitude just set. Set the Horizontal
SCALE
as
Table 18: Trigger settings for ≥ 4 GHz
■
CHECK that a stable trigger is obtained.
■
Read the following definition: A stable trigger is one where the
Trig'd
LED will remain constantly lighted.
■
Press the Trigger
Slope
button to select the positive slope.
■
Adjust the Trigger
Level
knob so that there is a stable trigger.
CHECK that the trigger is stable.
■
Press the Trigger
Slope
button to select the negative slope. Adjust the
Trigger
Level
knob so that there is a stable trigger.
■
CHECK that the trigger is stable.
■
Enter pass or fail in the test record.
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)