Horizontal and acquisition system specifications
Sample rate
Instrument
Maximum sample rate
Number of channels
MSO5204/B, DPO5204/B,
MSO5104/B, DPO5104/B
10 GS/s
1 or 2
5 GS/s
3 or 4
MSO5054/B, DPO5054/B,
MSO5034/B, DPO5034/B
5 GS/s
1 – 4
Record length, analog channels
Option
Number of
Channels
2 GHz
1 GHz
500 MHz
350 MHz
Standard (MSO/DPO5000 Series)
3 or 4
12.5 M
12.5 M
12.5 M
12.5 M
1 or 2
25 M
25 M
12.5 M
12.5 M
Option 2RL
(MSO/DPO5000 Series)
Standard
(MSO/DPO5000B Series)
3 or 4
25 M
25 M
25 M
25 M
1 or 2
50 M
50 M
25 M
25 M
Option 5RL
3 or 4
50 M
50 M
50 M
50 M
1 or 2
125 M
125 M
50 M
50 M
Option 10RL
3 or 4
125 M
125 M
125 M
125 M
1 or 2
250 M
250 M
125 M
125 M
Record length, digital channels
The record length ranges for a standard instrument are the same as for the analog channel
specification table (above).
The record length ranges for an instrument with an option are the same as for the analog channel
specifications, up to 40M.
Specifications (MSO/DPO5000/B Series)