Check sensitivity edge trigger DC coupled
Equipment required
Prerequisites
One leveled sine wave generator (Item 9)
Three precision 50 Ω coaxial cables (Item 21)
Two SMA female to BNC male adapters (Item
23)
Three SMA female-to-female adapters (Item 16)
One 10X attenuator (Item 1)
One power splitter (Item 11)
Male N-to-BNC adapter (Item 14)
One SMA adapter (Item 19)
One 5X attenuator (Item 2)
One 2X attenuator (Item 26)
< 4 GHz models, one 50 Ω termination (Item 3)
(See
on page 116.)
NOTE.
The sine wave generator output amplitude must be leveled to within
0.35 dB of the reference frequency (10 MHz) through the trigger frequency being
tested. Refer to the Sine Wave Generator Leveling Procedure if your sine wave
generator does not have automatic output amplitude leveling. (See
on page 281.)
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series
257