4.
Set the
Bandwidth
to
20 MHz
5.
Set the Horizontal
Scale
to
1 ms/division
.
6. Set the Acquisition mode to Average as follows:
a.
Push the front-panel
Acquire
button.
b.
In the Acquisition window, select the
Acquisition
tab at the left.
c.
Select
Average
Acquisition mode.
d.
Ensure that the
number of averages
is set to
16
.
7. Set the trigger source to AC line as follows:
a.
Push the front-panel Trigger
Menu
button.
b.
In the Trigger menu, under
Source
, click
Select
.
c.
Click
Line
to set the trigger source to AC line, and then click
OK
.
8.
Set the Vertical
Scale
to
1 mV
.
9. Select the mean measurement (if not already selected) as follows:
a.
Push the front-panel Wave Inspector
Measure
button.
b.
In the Measurements Setup window, go to the
Measurements
box. Click
the
More
button and then select
Mean
.
c.
View the
Mean
measurement value in the display.
10.
Enter the mean value as the test result in the test record.
11.
Repeat steps 8 through 10 for each vertical scale setting in the test record.
12.
Repeat steps 5 through 11 for each bandwidth setting in the test record
table.
13. Repeat the tests at 1 MΩ impedance as follows:
a.
Push the front-panel Vertical
V
menu button.
b.
Set the
Termination
to
1M Ω
.
c.
14. Repeat the procedure for all remaining channels as follows:
a.
Push the front-panel button to deselect the channel that you have already
tested.
b.
Push the front-panel button to select the next channel to be tested.
c.
Move the 50 Ω terminator to the channel input to be tested.
d.
Starting from step 3, repeat the procedure until all channels have been
tested.
15.
Press the
Menu Off
button.
Performance verification (MSO/DPO5000/B series)
320
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series