Delta Time Measurement Accuracy
Sample rate = 2.5 GS/s, 4 ns/Div, Source freq = 240 MHz
MSO/DPO V/Div
Source V
pp
Test result
High limit
1 mV
8 mV
NA
5 mV
40 mV
234 ps
100 mV
800 mV
233 ps
500 mV
4 V
233 ps
1 V
4 V
237 ps
Sample rate = 2.5 GS/s, 40 ns/Div, Source freq = 24 MHz
1 mV
8 mV
735 ps
5 mV
40 mV
422 ps
100 mV
800 mV
257 ps
500 mV
4 V
254 ps
1 V
4 V
581 ps
Sample rate = 2.5 GS/s, 400 ns/Div, Source freq = 2.4 MHz
1 mV
8 mV
6.99 ns
5 mV
40 mV
3.54 ns
100 mV
800 mV
2.72 ns
500 mV
4 V
2.69 ns
1 V
4 V
5.34 ns
Sample rate = 250 MS/s, 4 μs/Div, Source freq = 240 kHz
1 mV
8 mV
69.9 ns
5 mV
40 mV
35.4 ns
100 mV
800 mV
27.2 ns
500 mV
4 V
26.9 ns
1 V
4 V
53.4 ns
Sample rate = 25 MS/s, 40 μs/Div, Source freq = 24 kHz
1 mV
8 mV
699 ns
5 mV
40 mV
354 ns
100 mV
800 mV
272 ns
500 mV
4 V
269 ns
1 V
4 V
534 ns
Sample rate = 2.5 MS/s, 400 μs/Div, Source freq = 2.4 kHz
1 mV
8 mV
6.99 μs
5 mV
40 mV
3.54 μs
100 mV
800 mV
2.72 μs
500 mV
4 V
2.69 μs
1 V
4 V
5.34 μs
MSO/DPO5054/B and MSO/DPO5034/B Channel 4
Performance verification (MSO/DPO5000/B series)
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series
313