Click the
Trig After Time
button under A Then B.
Click the
B Trig Level
control in the control window.
Set the front-panel trigger mode to
Norm
.
b. Confirm that the following statements are true
:
■
The trigger-level readout for the B trigger system changes as you
turn the lower multipurpose knob.
■
As you rotate the lower multipurpose knob, the square-wave signal
can become triggered and untriggered. (Leave the signal triggered.)
c. Verify the delayed trigger counter:
■
Double click the
Trig Delay
control to pop up a numeric keypad for
that control.
■
Click the keypad to enter a trigger delay time of
1 second
(click
1 and None) and then click
Enter
.
■
Verify that the trigger
Ready
indicator on the front panel flashes
about once every second as the waveform is updated on-screen.
7. Remove the test hookup:
Disconnect the test hookup from the instrument.
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)