Verify the digital channels (MSO70000C/DX series only).
Equipment required
Prerequisites
MSO70000C/DX Series models: One precision 50 Ω
coaxial cable (item 4)
MSO70000C/DX Series models: One probe tip adapter
(item 30)
MSO70000C/DX Series models: One SMA adapter
(item 17)
MSO70000C/DX Series models: One logic probe
(item 31)
None
1. Hook up the signal source:
Connect the probe compensation or fast edge
output to the D0 input as shown in the following figure.
Figure 7: Setup for the digital channels test
2. Set up the instrument:
Push the front panel
Default Setup
button.
3.
Set the Horizontal Scale to
1 ms/div
.
4. Initialize the instrument:
Pull down the
Digital
menu and select
Digital
Setup
. This displays the instrument Digital Setup control window.
5.
Press the
D0
button.
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)