Edge trigger sensitivity not DC
coupled, typical
All sources, positive or negative edge, for vertical scale settings ≥10 mV/div and ≤1 V/div
≥4 GHz models
Trigger Coupling
Sensitivity
NOISE REJ
15%FS from DC to 50 MHz
35%FS at 3 GHz
50%FS at 5 GHz
AC
Same as DC-coupled limits for frequencies >
100 Hz, attenuates signals <100 Hz
HF REJ
Same as DC-coupled limits for frequencies <
20 kHz, attenuates signals > 20 kHz
LF REJ
Same as DC-coupled limits for frequencies >
200 kHz, attenuates signals < 200 kHz
<4 GHz models
Trigger coupling
Sensitivity
NOISE REJ
3 times the DC-coupled limits
AC
Same as DC-coupled limits for frequencies >
100 Hz, attenuates signals <100 Hz
HF REJ
Same as DC-coupled limits for frequencies <
20 kHz, attenuates signals > 20 kHz
LF REJ
Same as DC-coupled limits for frequencies >
200 kHz, attenuates signals < 200 kHz
Trigger level or threshold range
>20 GHz models
Trigger Source
Range
Any channel
±100% FS from center of screen
Auxiliary input
±4.0 V
Line
0 V, Not settable
≥4 GHz, ≤20 GHz models
Trigger Source
Range
Any channel
±120% FS from center of screen
Auxiliary input
±4.0 V
Line
0 V, Not settable
<4 GHz models
Trigger Source
Range
Any channel
±12 divisions from center of screen
Auxiliary input
±5 V
Line
Not settable
Trigger specifications (cont.)