8. Record the peak-to-peak measurement as follows:
a.
Push the front-panel Wave Inspector
Measure
button.
b.
In the
Measurements
box, click
Pk-Pk
.
c.
This will provide a mean V
p-p
of the signal. Call this value
V
in-pp
.
d.
Enter this value in the test record.
9.
Set the Horizontal
Scale
to
5 ns/division
.
10.
Adjust the signal source to the maximum bandwidth frequency for the
bandwidth and model desired, as shown in the following worksheet.
11. Record the peak-to-peak measurement as follows:
a.
With the
Pk-Pk
measurement still displayed from step 8, view the
measurement at the new frequency.
b.
This will provide a mean V
p-p
of the signal. Call this value
V
bw-pp
.
Performance verification (MSO/DPO5000/B series)
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series
327