bandwidth test hookup, < 3.5 GHz models
of the generator to a reference frequency of 50 MHz or less.
Figure 20: Analog bandwidth test hookup, < 3.5 GHz models
2. Confirm the input channels are within limits for analog bandwidth:
Do the
following substeps - test Ch 1 first,
skipping substeps
already set up for testing from step
a. Select an unchecked channel:
■
From the button bar, touch
Measure
and then
Clear All
to remove
the previous measurement.
■
Press the Vertical button of the channel just confirmed to remove the
channel from the display.
■
Press the front-panel Vertical button that corresponds to the channel
you are to confirm.
■
Touch
Vertical
, select
Vertical Setup
, and then touch Termination
50 Ω
.
■
Move the leveled output of the sine wave generator to the channel
you selected.
b. Match the trigger source to the channel selected:
Press the Trigger
Source
button until the source that corresponds to the channel you are to
confirm is on.
Table 12: Analog bandwidth, < 3.5 GHz models
Vertical scale
Reference
amplitude
Horizontal scale Test frequency
-3 dB
DPO7254C
DPO7104C
DPO7054C
5 mV
30 mV
1 ns
2.0 GHz
1 GHz
500 MHz
≥21.2 mV
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)