Scale setting
Position setting
(Divs)
Offset setting
4
Generator setting
Accuracy limits
50 mV
-5
+0.25 V
+650 mV
+636.75 mV to +663.25 mV
+5
-0.25 V
-650 mV
-663.25 mV to -636.75 mV
100 mV
-5
+2.0 V
+2.8 V
+2.764 V to 2.836 V
+5
-2.0 V
-2.8 V
-2.836 V to -2.764 V
200 mV
-5
+1.25 V
+2.75 V
+2.697 V to 2.803 V
+5
-1.25 V
-2.75 V
-2.803 V to -2.697 V
500 mV
-5
0 V
+3.5 V
+3.394 V to 3.606 V
+5
0 V
-3.5 V
-3.606 V to -3.394 V
< 4 GHz models, 50 Ω and 1 MΩ
1 mV
-5
+0.5 V
+507 mV
+504.24 mV to +509.76 mV
+5
-0.5 V
-507 mV
-509.76 mV to -504.24 mV
2 mV
-5
+0.5 V
+516 mV
+512.97 mV to +519.03 mV
+5
-0.5 V
-516 mV
-519.03 mV to -512.97 mV
5 mV
-5
+0.5 V
+540 mV
+536.26 mV to +543.74 mV
+5
-0.5 V
-540 mV
-543.74 mV to -536.26 mV
10 mV
-5
+0.5 V
+580 mV
+574.19 mV to +585.81 mV
+5
-0.5 V
-580 mV
-585.81 mV to -574.19 mV
20 mV
-5
+0.5 V
+660 mV
+651.62 mV to +668.38 mV
+5
-0.5 V
-660 mV
-668.38 mV to -651.62 mV
50 mV
-5
+0.5 V
+900 mV
+883.81 mV to +916.19 mV
+5
-0.5 V
-900 mV
-916.19 mV to -883.81 mV
90 mV
-5
+0.5 V
+1.22 V
+1.192 V to +1.248 V
+5
-0.5 V
-1.22 V
-1.248 V to -1.192 V
100 mV
0
+4.5 V
+4.8 V
+4.746 V to +4.854 V
0
-4.5 V
-4.8 V
-4.854 V to -4.746 V
200 mV
-2
+4.6 V
+4.8 V
+4.725 V to +4.875 V
+2
-4.6 V
-4.8 V
-4.875 V to -4.725 V
500 mV
-2
+5.0 V
+4.5 V
+4.345 V to +4.655 V
+2
-5.0 V
-4.5 V
-4.655 V to -4.345 V
1 V 50 Ω
0
+2.0 V
+5.0 V
+4.742 V to +5.258 V
0
-2.0 V
-5.0 V
-5.258 V to -4.742 V
1 V 1 MΩ
0
+96.0 V
+99.0 V
+98.390 V to +99.610 V
0
-96.0 V
-99.0 V
-99.610 V to -98.390 V
d. Display the test signal:
■
From the button bar touch
Vertical
and touch
Position
.
4
Set as precisely as the instrument's offset resolution permits.