DC Gain Accuracy
Performance checks Bandwidth
Vertical scale
Low limit
Test result
High limit
Channel 4 DC Gain
Accuracy, 0 V offset,
0 V vertical position,
50 Ω
20 MHz
2 mV/div
-1.5%
1.5%
5 mV/div
-1.5%
1.5%
10 mV/div
-1.5%
1.5%
20 mV/div
-1.5%
1.5%
50 mV/div
-1.5%
1.5%
100 mV/div
-1.5%
1.5%
200 mV/div
-1.5%
1.5%
500 mV/div
-1.5%
1.5%
1 V/div
-1.5%
1.5%
250 MHz
20 mV/div
-1.5%
1.5%
FULL
20 mV/div
-1.5%
1.5%
Channel 4 DC Gain
Accuracy, 0 V offset,
0 V vertical position,
1 MΩ
20 MHz
2 mV/div
-1.5%
1.5%
5 mV/div
-1.5%
1.5%
10 mV/div
-1.5%
1.5%
20 mV/div
-1.5%
1.5%
50 mV/div
-1.5%
1.5%
100 mV/div
-1.5%
1.5%
200 mV/div
-1.5%
1.5%
500 mV/div
-1.5%
1.5%
1 V/div
-1.5%
1.5%
250 MHz
20 mV/div
-1.5%
1.5%
FULL
20 mV/div
-1.5%
1.5%
DC Offset Accuracy
Performance checks Vertical scale
Vertical offset
1
Low limit
Test result
High limit
All models:
Channel 1 DC Offset
Accuracy, 20 MHz
BW, 50 Ω
1 mV/div
900 mV
895.3 mV
904.7 mV
1 mV/div
-900 mV
-904.7 mV
-895.3 mV
100 mV/div
5.0 V
4.965 V
5.035 V
100 mV/div
-5.0 V
-5.035 V
-4.965 V
1
Use this value for both the calibrator output and the oscilloscope offset setting.
Performance verification (MSO/DPO5000/B series)
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series
293
Summary of Contents for DPO5000 Series
Page 2: ......
Page 6: ......
Page 352: ...Index 340 MSO70000C DX DPO70000C DX DPO7000C MSO5000 B DPO5000 B Series...