6.
Double click
Global Threshold
. Use the keypad to enter
200 mV
.
Press the
Apply
button.
7.
Pull down the
Vertical
menu and select
Vertical Setup
.
8.
Press Digital input
iCapture
.
9.
From the Digital Source list select
D0
.
10.
Pull down the
Trig
menu and select
A Event (Main) Trigger Setup
.
11.
From the Source list select
D0
.
12.
Verify that the both the iCapture and the digital channel are triggered.
13.
Rotate the Vertical
SCALE
knob counter clockwise.
14.
Verify that the displayed amplitude of channel 1 is reduced.
15.
Rotate the Vertical
SCALE
knob clockwise, back to its original setting.