6. Disconnect the hookup:
a.
Disconnect all test equipment from the instrument.
b. Set the instrument controls:
■
From menu mode, touch
Utilities
and select
External Signals
.
■
Touch the
Internal
button to select the internal reference.
c. Perform a signal path compensation:
■
Touch
Utilities
and select
Instrument Calibration
.
■
Touch
Run SPC
(
Calibrate
on some instruments) and wait for the
signal path compensation to finish.
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series