3. Record the reference level:
■
Disconnect the sine wave generator from the instrument.
■
Connect the sine wave generator to the power sensor.
■
Note the level meter reading.
4. Set the generator to the new frequency and reference level:
■
Change the sine wave generator to the desired new frequency.
■
Input the correction factor and/or the new frequency into the level meter.
■
Adjust the sine wave generator amplitude until the level meter again
reads the value noted in step 3. The signal amplitude is now correctly set
for the new frequency.
■
Disconnect the sine wave generator from the power sensor.
■
Connect the sine wave generator to the instrument.
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
284
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series