Pulse response peak detect or
envelope mode <4 GHz models
This instrument uses analog peak detection for pulse capture in Peak Detect or Envelope mode at
sample rates of 125 MS per second and slower. At faster sample rates the instrument uses
conventional sampling.
The minimum single pulse widths for 50% or greater amplitude capture for pulses greater than
2 divisions in magnitude.
Sample rate setting
Minimum pulse width
≤ 10 GS/s
1 ÷ (sample rate) or 100 ps
40 MS/s
1 ÷ (sample rate) or 25 ps
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Vertical system analog channels