![Stryker IsoTour 2872 Operation Manual Download Page 18](http://html.mh-extra.com/html/stryker/isotour-2872/isotour-2872_operation-manual_1385944018.webp)
N
No
otte
e -- The maximum drying temperature should not exceed 60° C (140° F).
N
No
otte
e -- Stryker recommends that you do not use softening agents in the washing or drying machines. These agents
leave a residue that change the breathability properties of the fabric.
2. Dry the top cover for 90 minutes or until the top cover is dry to the touch.
C
Ca
arre
e a
an
nd
d m
ma
aiin
ntte
en
na
an
ncce
e
W
WA
AR
RN
NIIN
NG
G
•
Do not wash the internal components of this support surface. Discard the support surface if contamination is found
inside.
•
Do not immerse the product in cleaning or disinfectant solutions.
•
Do not allow liquid to pool on the product.
•
Always inspect support surface covers (top and bottom) for tears, punctures, excessive wear, and misaligned zippers
before each use. If compromised immediately remove the support surface from service.
•
Always make sure that you wipe each product with clean water and dry each product after you use cleaning agents.
Some cleaning agents are corrosive in nature and may cause damage to the product. Failure to follow these care and
maintenance instructions may void your warranty.
C
CA
AU
UT
TIIO
ON
N
•
Do not allow liquid to seep into the zipper area or watershed cover barrier when you clean the underside of the support
surface. Fluids allowed to come in contact with the zipper may leak into the support surface core.
•
Do not iron, dry-clean, or tumble dry the support surface covers.
•
Do not power wash the support surface as this may damage the product.
•
Always dry the support surface covers before you store, add linens, or place a patient on the surface. A dry product
helps to prevent impaired performance.
•
Do not over expose the covers to higher concentration chemical solutions as these may degrade the covers.
•
Failure to follow manufacturing instructions may also affect useful life of the support surface cover.
The support surface cover is resistant to the following chemical solutions:
•
Quaternaries (active ingredient - ammonium chloride) that contain less than 3% glycol ether
•
Phenolic solution (Matar)
•
Chlorinated bleach solution (6500 ppm)
•
Accelerated hydrogen peroxide
•
Potassium peroxymonosulfate
•
70% isopropyl alcohol
Follow hospital protocol for support surface care between patients, to avoid the risk of cross-contamination and infection.
C
Ca
arre
e o
off b
bo
otttto
om
m cco
ovve
err
Always follow your hospital protocol for cleaning and disinfecting. This is a guide to help show you how to access the
bottom surface cover for cleaning.
N
No
otte
e -- This product is compatible with different bed frames. The bed features listed here may not be available for all
compatible beds frames.
1. Remove the head board and foot board from the bed and set aside.
2. Lower the siderails (Figure 4).
EN
14
2872-009-005 Rev B.1
Summary of Contents for IsoTour 2872
Page 2: ......
Page 24: ......
Page 26: ......
Page 48: ......
Page 50: ......
Page 72: ......
Page 74: ......
Page 98: ......
Page 100: ......
Page 124: ......
Page 126: ......
Page 150: ......
Page 152: ......
Page 174: ......
Page 176: ......
Page 198: ......
Page 200: ......
Page 224: ......
Page 226: ......
Page 248: ......
Page 250: ......
Page 272: ......
Page 274: ......
Page 298: ......
Page 300: ......
Page 322: ......
Page 324: ......
Page 348: ......
Page 372: ......
Page 394: ......
Page 396: ......
Page 420: ......
Page 422: ......
Page 446: ......
Page 448: ......
Page 472: ......
Page 474: ......
Page 498: ......
Page 522: ......
Page 544: ......
Page 546: ......
Page 568: ......
Page 570: ......
Page 579: ...序 序列 列号 号 A 生 生产 产日 日期 期 生产日期是序列号的前四位数字 2872 009 005 Rev B 1 7 ZH ...
Page 591: ......