Operate
6.2 Parameter description
SITRANS P500 with HART
98
Operating Instructions, 12/2014, A5E02344528-09
"2.1.8.4 Curve Activation"
The description of this parameter and the following ones can be found under:
"1.1.10.4 Level Curve" (Page 80)
"2.1.8.66 Density Unit"
Factory setting:
g/cm
3
Adjustment range:
g/cm
3
, kg/m
3
, lb/Gal, lb/ft
3
, g/ml, kg/l, g/l, lb/in
3
, ug/l, ug/m
3
Purpose:
Selection of the unit for the density parameter
Description:
This parameter allows you to select the unit for the measured density value.
"2.1.8.67 Density"
Factory setting:
1 g/cm
3
Adjustment range:
Freely selectable numeric value
Purpose:
Setting the value for the density
Description:
This parameter allows you to set the value for the density of the medium,
which is used for conversion between volume and mass.
Adjust the settings for the measurement range when you change the values
for density!
"2.1.9 User"
Note
This menu command is visible only if "user" has been selected under "2.1.1 PV Select"
(Page 83).
See also
"User" measurement type (Page 153)
"2.1.9.1 User Curve Unit"
Factory setting:
AAAAA
Setting range:
5 characters
Purpose:
Enter a user-specific unit
Description:
This parameter allows you to set the unit of the user-specific measured
variable. The parameters associated with this unit are automatically con-
verted into the new unit.
The user-specific unit consists of up to 5 characters, freely selectable.