Alarm, error and system messages
9.1 Overview of messages and symbols
SITRANS P500 with HART
186
Operating Instructions, 12/2014, A5E02344528-09
The sequence of the symbols corresponds to the priority of the messages, beginning with
the most serious.
Icon
Category
1)
Definition
Simulation
Simulation or substitute value
Maintenance alarm
The device outputs fault current.
Service the device immediately.
Maintenance demanded The service interval that you set has been exceeded.
Service the device without delay.
- or -
The pressure measurement value is unreliable.
Process value alarm
The device outputs fault current or is at the limit of the saturation range.
Process value warning
There is a problem with one or more process values. Thus the device is still measur-
ing process values, but these may be unreliable.
Example: A process value exceeds the device specification.
1)
The category of the messages is shown in the bottom line of the display.