![Siemens SITRANS LR200 7ML5422 Operating Instructions Manual Download Page 87](http://html.mh-extra.com/html/siemens/sitrans-lr200-7ml5422/sitrans-lr200-7ml5422_operating-instructions-manual_384663087.webp)
SITRANS LR200 with mA/HART
Operating Instructions, 05/2019, A5E32337676-AE
87
TVT
Modify the TVT to screen out false echoes [see Auto False Echo Suppression (2.8.7.1.)
(Page 115)].
Navigate to Configure/Setup > Setup > Signal Processing and click on TVT. Click on one of
the two tabs to access the parameters listed.
TVT Setup [see TVT setup (2.8.7.)]
●
Auto False Echo Suppression
●
Auto False Echo Suppression Range
●
Hover Level
●
Shaper Mode
TVT Shaper
●
Shaper breakpoints 1 to 40. (Turn TVT Setup/Shaper Mode on to activate.)
Manual TVT Curve
Displays the effects of the TVT shaper modifications. Navigate to Configure/Setup > Setup >
Signal Processing and click on Manual TVT Curve.
Echo Profile
●
Navigate to Configure/Setup > Setup > Signal Processing and click on Echo Profile.
●
Select Standard operation for faster display.