Configuring a stillpipe application
●
Launch the Quick Start wizard and follow steps 1 to 2.
●
In step 2, select Application Type – Level in a vessel with stillpipe, Material – Liquid Low
dK, and set the stillpipe diameter as desired.
●
The wizard updates the propagation factor according to the pipe diameter, and enables
the CLEF algorithm for low dK liquids [see Position detect (2.8.4.2.) (Page 112) for more
detail].
●
Continue through steps 3 to 5 then click on FINISH AND DOWNLOAD to save settings
offline and transfer them to the device.
6.3.1.4
Changing parameter settings using SIMATIC PDM
Note
•
For a complete list of parameters, see Parameter Reference (Page 97).
•
Clicking on Cancel during an upload from device to SIMATIC PDM will result in some
parameters being updated.
Many parameters are accessed via pull-down menus in PDM. See Parameters accessed via
pull-down menus (Page 64) for others.
1.
Launch SIMATIC PDM, connect to device, and upload data from device.
2.
Adjust parameter values in the parameter value field then Enter. The status fields read
Changed.
3.
Open the Device menu, click on Download to device, then use File - Save to save
settings offline. The status fields are cleared.