![Siemens SITRANS LR200 7ML5422 Operating Instructions Manual Download Page 135](http://html.mh-extra.com/html/siemens/sitrans-lr200-7ml5422/sitrans-lr200-7ml5422_operating-instructions-manual_384663135.webp)
SITRANS LR200 with mA/HART
Operating Instructions, 05/2019, A5E32337676-AE
135
7.6.2.2
Pin to unlock (6.2.2.)
Description
Stores the value to be entered in Write Protect (6.2.1.) (Page 134) to unlock
programming. If Write Protect (6.2.1.) (Page 134) is set to a different value, PIN
to Unlock (6.2.2.) does not display the Unlock value. Valid only for operation via
the handheld programmer.
Handheld pro-
grammer values
Range: 0 to 99999
Value when shipped: 1954.
Not restored by a reset to Factory Defaults.
----
Display when Lock is on
Note
Do not lose your Unlock Value. It cannot be displayed once Write Protect (6.2.1.) (Page 134)
has been set to a different value.
7.7
Language (7.)
Description
Selects the language to be used on the LCD.
Options
English (factory setting)
Deutsch
Français
Español