SITRANS LR200 with mA/HART
60
Operating Instructions, 05/2019, A5E32337676-AE
Step 5 - Summary
Check parameter settings, and click on BACK to return and revise values, FINISH to save
settings offline, or FINISH AND DOWNLOAD to save settings offline and transfer them to the
device.
The message Quick Start was successful will appear. Click on OK.
Linearization
You can use the linearization feature to define a more complex vessel shape and enter up to
32 level breakpoints where the corresponding volume is known. The values corresponding to
100% and 0% levels must be entered. The breakpoints can be ordered from top to bottom,
or the reverse.
Example (values for example purposes only)
Break-
point no.
Level
value (m)
Volume
value (l)
1
0
0
2
5
500
3
9
3000
4
19.5
8000