3.3
Article Number 2. 540 028
SCHILLER AG 2002
AT-102 Service Handbook
Section 3
Fault Finding
Fault Finding Chart
IS THE GREEN MAINS
LED LIT?
NO
CONNECT MAINS
TO THE UNIT
AND SWITCH ON
FUSE OK?
NO
YES
FUSE OK?
NO
DISCONNECT THE MAINS
SUPPLY AND CHECK THE
MAINS FUSES (IN THE
MAINS MODULE)
INVESTIGATE THE REASON
FOR THE RUPTURED FUSE
YES
AT-102 Initial Fault Diagnosis Chart (Sheet 1)
CHECK THE BATTERY, THE
BATTERY LEADS AND THE
BATTERY CONNECTORS
FOR SHORT CIRCUITS.
REPLACE FUSE AND
SWITCH ON. IF THE FUSE
IS STILL RUPTURING
REPLACE THE MAINBOARD
GO TO FAULT DIAGNOSIS
(SHEET A - POWER PRO-
BLEMS)
NOTE:
The removal and replacement instructions and the location of all boards, cable
assemblies and connectors are given in Chapter 4.
PRESS THE
<ON> KEY
ENSURE THAT THE
VOLTAGE SELECTOR IS
CORRECTLY SET.
REPLACE THE MAINS
FUSE AND SWITCH THE
MAINS ON
DOES THE LCD LIGHT?
NO
YES
SHEET
TWO
GO TO FAULT DIAGNOSIS
(SHEET B - POWER PRO-
BLEMS)
GO TO FAULT DIAGNOSIS
(SHEET C - GENERAL PRO-
BLEMS)
A
YES
Summary of Contents for AT-102
Page 18: ...xviii SCHILLER AT 102 ECG Recorder Service Handbook ...
Page 26: ...1 8 AT 102 Service Handbook Introduction Keypad 4 7 8 9 10 11 12 13 14 15 16 1 2 3 5 6 17 ...
Page 54: ...1 36 AT 102 Service Handbook ...
Page 62: ...2 8 AT 102 Service Handbook ...
Page 78: ...3 16 AT 102 Service Handbook ...
Page 98: ...5 2 AT 102 Service Handbook ...
Page 114: ...5 18 AT 102 Service Handbook ...
Page 118: ...6 4 AT 102 Service Handbook ...
Page 120: ...7 2 AT 102 Service Handbook ...
Page 126: ...7 8 AT 102 Service Handbook ...
Page 134: ...Annex A 8 AT 102 Service Handbook ...
Page 136: ...Index 1 Article Number 2 540 028 SCHILLER AG 2002 AT 102 Service Handbook Index Index ...
Page 140: ...Index 5 Article Number 2 540 028 SCHILLER AG 2002 AT 102 Service Handbook Index ...