SCHILLER AG 2002
AT-102 Service Handbook
Section 3
Fault Finding
AT-102 Fault Diagnosis Chart (Sheet H - Communication (RS)
Problems)
CONFIRM SETTINGS IN
THE RS CHECK
PROCEDURE DETAILED IN
CHAPTER 5.
NO
SETTING CORRECT?
YES
POSSIBLE OPERATOR
ERROR.
CHECK ALL CABLE
ASSEMBLIES. REPLACE IF
NECESSARY
1. CHECK THE +12V AND -12V
SUPPLY VOLTAGES
2. CHANGE THE MAIN BOARD
ENTER SETTINSG MENU
CHECK THAT ALL SETTINGS(
LINE/MODEM, RECORD/BLOCK,
PARITY, STOP BIT AND THE BAUD
RATE) ARE CORRECT FOR THE
APPLICATION AND ARE THE
SAME IN BOTH DEVICES.