1.28
AT-102 Service Handbook
System Settings
Communication
Baudrate
Select a Baud rate between 115200 and 9600 Baud, according to the
modem/computer used. Most computers can connect at 115200 Baud
and the standard modem speed is 57600 Baud. If problems are
experienced during transmission reduce the Baud rate.
Mode
Select between line (computer connected directly to the RS-232
interface), or modem (for transmission over the phone network)
Phone No.
Enter the telephone number preceded by `T` or `P` (tone or pulse).
A comma `,` gives a one second pause in dialling - this may be
necessary for example, if an outside line is required.
Modem Init.
Enter the modem initialisation codes. Full details will be found in the
user guide for your modem. However, the modem initialisation must
contain at the minimum, the following commands with the prefix `AT`.
`Q0`- modem sends response
`V0`- numerical response codes
`E0`- no command echo
The standard modem initialisation code is: ATB0L1V0Q0E0S0=0
If in doubt about any of these settings, please contact your phone company and/or modem
supplier.
Summary of Contents for AT-102
Page 18: ...xviii SCHILLER AT 102 ECG Recorder Service Handbook ...
Page 26: ...1 8 AT 102 Service Handbook Introduction Keypad 4 7 8 9 10 11 12 13 14 15 16 1 2 3 5 6 17 ...
Page 54: ...1 36 AT 102 Service Handbook ...
Page 62: ...2 8 AT 102 Service Handbook ...
Page 78: ...3 16 AT 102 Service Handbook ...
Page 98: ...5 2 AT 102 Service Handbook ...
Page 114: ...5 18 AT 102 Service Handbook ...
Page 118: ...6 4 AT 102 Service Handbook ...
Page 120: ...7 2 AT 102 Service Handbook ...
Page 126: ...7 8 AT 102 Service Handbook ...
Page 134: ...Annex A 8 AT 102 Service Handbook ...
Page 136: ...Index 1 Article Number 2 540 028 SCHILLER AG 2002 AT 102 Service Handbook Index Index ...
Page 140: ...Index 5 Article Number 2 540 028 SCHILLER AG 2002 AT 102 Service Handbook Index ...