4.14
AT-102 Service Handbook
Main Board MK 18 - 1
Board Replacement
To replace the Main board MK 18-1 proceed as follows:
1.
Position the board and secure at the 12 fixing points
2.
Reconnect all connecters
P2 (under)
internal battery
P9 (under)
Video (VGA)
P15 (under)
Ext Printer
P14 (under)
RS-232
P1 (under)
Patient
P16 (under)
Comm Module
(not used)
P8
LCD (Data)
P10
Keyboard
P7
Thermal Printer (Motor)
P12
Paper Mark Detector
P11
Thermal Printer (Data)
P5 Blood
pressure Module
(not used)
P13
Paper Tray
Motor