AT-102 ECG Service Handbook
Section 1
Operating Elements
Introduction
Items 4 to 10 are specific for ECG acquisition only:
4.
Current Heart Rate (averaged over 4 beats and refreshed every 2 seconds). The HR is
also given on a manual printout.
Note that with an auto mode printout the HR is
averaged over the full 10 seconds of the recording.
5.
Electrode connections - when a lead flashes it indicates that the electrode resistance
is too high. The electrode(s) must be reapplied.
6.
Sensitivity - 5, 10 or 20 mm/mV. Change the sensitivity with the keys 3 (auto), 4, 5
and 6. An `A` in this box indicates that automatic sensitivity is selected (auto mode
printout only).
7.
Speed - 25 or 50 mm/s. Change the speed with the keys 8 and 9.
8.
Lead indication (leads currently displayed on the screen). Change the lead group with
the < and > keys on the keypad.
9.
Myogram Filter indication - 'Filter ON' or 'Filter OFF'. The filter is applied with the filter
key.
Note: the frequency of the filter cutoff is defined in Section 4 Setup.
10.
Area for system messages or instructions.
Summary of Contents for AT-102
Page 18: ...xviii SCHILLER AT 102 ECG Recorder Service Handbook ...
Page 26: ...1 8 AT 102 Service Handbook Introduction Keypad 4 7 8 9 10 11 12 13 14 15 16 1 2 3 5 6 17 ...
Page 54: ...1 36 AT 102 Service Handbook ...
Page 62: ...2 8 AT 102 Service Handbook ...
Page 78: ...3 16 AT 102 Service Handbook ...
Page 98: ...5 2 AT 102 Service Handbook ...
Page 114: ...5 18 AT 102 Service Handbook ...
Page 118: ...6 4 AT 102 Service Handbook ...
Page 120: ...7 2 AT 102 Service Handbook ...
Page 126: ...7 8 AT 102 Service Handbook ...
Page 134: ...Annex A 8 AT 102 Service Handbook ...
Page 136: ...Index 1 Article Number 2 540 028 SCHILLER AG 2002 AT 102 Service Handbook Index Index ...
Page 140: ...Index 5 Article Number 2 540 028 SCHILLER AG 2002 AT 102 Service Handbook Index ...