5.16
AT-102 Service Handbook
Upgrading the Unit / Updating the
Software
Updating the System Software
The AT-102 software is updated using a PC program called SWUP. To update the software,
the RS-232 connector on the side of the AT-102 is connected to a PC using a RS-232 to
RS-232 cable assembly.
Prerequisites
RS-232 cable assembly P/N 2.310 159.
SWUP Programme for Windows
TM
( this can be downloaded from the SCHILLER
Mailbox)
Procedure
1.
Using the RS-232 cable assembly P/N 2.310 159, connect the AT-102 (RS-232
connector on the side of the unit) to any free COM port on your PC.
2.
On the AT-102, enter the TEST and INFO screen
SETUP > SYSTEM SETTINGS >TEST AND INFO
3.
Select SOFTWARE. The following is displayed:
4.
Select UPDATE
Summary of Contents for AT-102