4.16
AT-102 Service Handbook
Battery Pack
Battery Pack Removal
WARNING:
The Warnings and Cautions at the beginning of the chapter must be
observed.
Ensure that the mains cable is removed before commencing
To remove the Battery Pack proceed as follows:
1.
Remove the top assembly as detailed previously.
2.
Disconnect the two bayonet connectors for the Main board (P2).
3.
Gently return the top assembly in position, and firmly holding both parts, turn the unit
up-side-down.
4.
Unscrew the battery compartment cover plate retaining screws and remove the
battery.
Checks and Tests After Battery Replacement
Program all static settings which will have been lost when the battery was disconnected
including date and time.