5.8
AT-102 Service Handbook
Paper Mark Detector Check
This procedure is a check only. There are no presets that can be adjusted. If this check
fails the main board MK 18-1 must be replaced.
Tools, Equipment and Material
Digital voltmeter
Small flat bladed screwdriver
70% alcohol solution and clean lint free cloth
Procedure
To check the paper-mark Schmitt trigger sensitivity proceed as follows:
WARNING
Mains power is potentially lethal - disconnect the mains before disas-
sembling the unit and remove the mains cable.
Paper
Mark
detector
cable
Paper
Mark
opto
sensor