AT-102 Service Handbook
Section 4
Module Removal and Replacement
Opening and Closing the Case
Top Assembly Removal
The Top Assembly is secured to the base assembly with seven recessed screws. Access
to the screws is gained from the underside of the unit. To remove the Top Assembly,
proceed as follows
WARNING
Ensure that the mains cable is removed.
1.
Turn the unit up-side-down and rest on a soft antistatic cloth.
2.
Unscrew and remove the six countersunk retaining screws and washers situated in
the extreme corners and edges of the unit.
4.
Grasping the top and bottom of the unit to ensure that the two assemblies cannot
part, carefully return the unit to the standing position.
5.
Gently lift the Top Assembly sufficiently to gain access to the interconnecting
cables. Disconnect the cable assembly between the main board MK 18-1 and the
keyboard and the ribbon and dual-wire cable assemblies between the power supply
and the LCD screen board.
6.
Gently lift the Top Assembly away from the Base Assembly and place on a soft
cloth.
Summary of Contents for AT-102
Page 18: ...xviii SCHILLER AT 102 ECG Recorder Service Handbook ...
Page 26: ...1 8 AT 102 Service Handbook Introduction Keypad 4 7 8 9 10 11 12 13 14 15 16 1 2 3 5 6 17 ...
Page 54: ...1 36 AT 102 Service Handbook ...
Page 62: ...2 8 AT 102 Service Handbook ...
Page 78: ...3 16 AT 102 Service Handbook ...
Page 98: ...5 2 AT 102 Service Handbook ...
Page 114: ...5 18 AT 102 Service Handbook ...
Page 118: ...6 4 AT 102 Service Handbook ...
Page 120: ...7 2 AT 102 Service Handbook ...
Page 126: ...7 8 AT 102 Service Handbook ...
Page 134: ...Annex A 8 AT 102 Service Handbook ...
Page 136: ...Index 1 Article Number 2 540 028 SCHILLER AG 2002 AT 102 Service Handbook Index Index ...
Page 140: ...Index 5 Article Number 2 540 028 SCHILLER AG 2002 AT 102 Service Handbook Index ...