3.15
Article Number 2. 540 028
SCHILLER AG 2002
AT-102 Service Handbook
Section 3
Fault Finding
RS-232 Interface (and Spiro Check)
There are no specific tests that can be performed on the RS-232 interface. However the RS-
232 (UART) settings can be checked to ensure correct protocol and speed. If a spiro sensor
is connected to the RS-232 connector, the sensor type can be displayed to ensure that it is
the correct type. The Comm test screen is entered for the settings screen:
(MENU) >SETUP > SYSTEM SETTING > TEST and INFO > COMM TEST
You are then prompted to select UART TEST and (only if a spiro sensor is connected to the
RS-232 interface), SPIRO SENSOR.
When UART TEST is selected the current Baud rate, RTS and CTS settings are displayed.
When SPIRO SENSOR is selected the sensor type is displayed (SP-250 or SP-260), and
the software version of the spirometry software.
Summary of Contents for AT-102
Page 18: ...xviii SCHILLER AT 102 ECG Recorder Service Handbook ...
Page 26: ...1 8 AT 102 Service Handbook Introduction Keypad 4 7 8 9 10 11 12 13 14 15 16 1 2 3 5 6 17 ...
Page 54: ...1 36 AT 102 Service Handbook ...
Page 62: ...2 8 AT 102 Service Handbook ...
Page 78: ...3 16 AT 102 Service Handbook ...
Page 98: ...5 2 AT 102 Service Handbook ...
Page 114: ...5 18 AT 102 Service Handbook ...
Page 118: ...6 4 AT 102 Service Handbook ...
Page 120: ...7 2 AT 102 Service Handbook ...
Page 126: ...7 8 AT 102 Service Handbook ...
Page 134: ...Annex A 8 AT 102 Service Handbook ...
Page 136: ...Index 1 Article Number 2 540 028 SCHILLER AG 2002 AT 102 Service Handbook Index Index ...
Page 140: ...Index 5 Article Number 2 540 028 SCHILLER AG 2002 AT 102 Service Handbook Index ...