3.5
Article Number 2. 540 028
SCHILLER AG 2002
AT-102 Service Handbook
Section 3
Fault Finding
AT-102 Fault Diagnosis Chart (Sheet A - Power Problems)
NO
YES
ARE ALL THE CABLE
ASSEMBLIES AND
CONNECTORS OK?
IS THE MAINS FUSE
OK?
NO
YES
SHORT CIRCUIT IN THE
MAINS TRANSFORMER.
REPLACE THE MAINS
TRANSFORMER
SHORT CIRCUIT IN THE
POWER SUPPLY. REPLACE
THE MAIN BOARD
IS THE MAINS FUSE
OK?
YES
DISCONNECT THE FLYING
LEADS FROM THE MAINS
TRANSFORMER TO THE
POWER SUPPLY. REPLACE
THE MAINS FUSE AND
SWITCH ON.
NO
REMOVE THE TOP ASSEMBLY .
VISUALLY INSPECTALL CABLE
ASSEMBLIES AND
CONNECTORS. ENSURE THAT
ALL CONNECTORS ARE SECURE.
LOOK IN THE AREA OF THE
TRANSFORMER AND MAINS
INPUT FOR SIGNS OF BURNING
OR SCORCHING.
SHORT CIRCUIT IN THE
MAINS ASSEMBLY.
REPLACE THE MAINS
ASSEMBLY
RECTIFY THE CAUSE OF
THE PROBLEM. REPLACE
FAULTY COMPONENT(S).
DISCONNECT THE
INTERNAL CABLES TO THE
MAINS ASSEMBLY. REPLACE
THE MAINS FUSE AND
SWITCH ON.
MAINS FUSE RUPTURED
Summary of Contents for AT-102
Page 18: ...xviii SCHILLER AT 102 ECG Recorder Service Handbook ...
Page 26: ...1 8 AT 102 Service Handbook Introduction Keypad 4 7 8 9 10 11 12 13 14 15 16 1 2 3 5 6 17 ...
Page 54: ...1 36 AT 102 Service Handbook ...
Page 62: ...2 8 AT 102 Service Handbook ...
Page 78: ...3 16 AT 102 Service Handbook ...
Page 98: ...5 2 AT 102 Service Handbook ...
Page 114: ...5 18 AT 102 Service Handbook ...
Page 118: ...6 4 AT 102 Service Handbook ...
Page 120: ...7 2 AT 102 Service Handbook ...
Page 126: ...7 8 AT 102 Service Handbook ...
Page 134: ...Annex A 8 AT 102 Service Handbook ...
Page 136: ...Index 1 Article Number 2 540 028 SCHILLER AG 2002 AT 102 Service Handbook Index Index ...
Page 140: ...Index 5 Article Number 2 540 028 SCHILLER AG 2002 AT 102 Service Handbook Index ...