1.15
Article Number 2. 540 028
SCHILLER AG 2002
AT-102 ECG Service Handbook
Section 1
Operating Elements
ECG Settings
Filters
There are five different filters which can be set individually as follows:
Baseline filter
The cutoff frequency of the filter is set on the top line. The cutoff can be 0.05Hz, 0.15Hz or
0.3Hz.
Note:
The set value is the lower limit of the frequency range and is normally set to
0.05 Hz. The settings 0.15 and 0.30 Hz should only be used when absolutely
necessary, as the possibility exists that they could affect the original ECG
signal, especially the ST segments.