AT-102 Service Handbook
AT-102 Fault Diagnosis Chart (Sheet G - Spirometry
Problems)
NO
YES
ENTER THE SPIROMETRY
MENU (<SPIRO>. ENSURE
THAT THE CALIBRATION IS
CORRECT AND THAT ALL
SETTINGS ARE CORRECT.
SEE AT-102 USER GUIDE
POSSIBLE OPERATOR
ERROR.
REPLACE THE
MOUTHPIECE (SP250)
REPLACE FILTER (SP-260)
CAN THE
CALIBRATION
PROCEDURE BE
CARRIED OUT?
CHANGE
1. MOUTHPICE
2. FLOW SENSOR.
3. MAIN BOARD
CHECK ALL SPIROMETRY
CABLING. ENSURE THAT ALL
CONNECTIONS ARE
SECURE.
CHECK SPIRO INTERFACE
SETTING DETAILED IN
CHAPTER 5. ENSURE
CORRECT SENSOR
SPECIFIED (SP-250 OR SP260)