5.10
AT-102 Service Handbook
ECG Amp2V, -2V and PWM
Ramp Time Adjustment
The ±2V voltage rails generated on the ECG Amplifier board are used as a reference by the
measurement and PWM circuits.
NOTE
The ±2V reference voltages, and the PWM ramp must both be adjusted at
the same time.
Tools, Equipment and Material
Digital voltmeter
Small flat bladed screwdriver
Procedure
WARNING
Mains power is potentially lethal - disconnect the mains before disas-
sembling the unit and remove the mains cable.
1.
Disassemble the unit as detailed in Chapter 4.
2.
Remove the Main board MK18-1 from the assembly and remove the EMC shield from
the board
3.
Replace the board in position
4.
Reconnect the keyboard to the control board and place the top assembly on its side
in front of the Base assembly .