3.9
Article Number 2. 540 028
SCHILLER AG 2002
AT-102 Service Handbook
Section 3
Fault Finding
AT-102 Fault Diagnosis Chart (Sheet F - Exercise Mode
Problems)
NO
YES
NB
ENSURE THAT THE
STRESS TEST PARAMETERS ARE
CORRECTLY SET (SEE CHAPTER 1
OR AT-102 OPERATING HANDBOOK)
NB
ENSURE THAT THE
STRESS TEST EQUIPMENT IS
FUNCTIONING CORRECTLY
IS A DIGITAL, RS
CONTROLLED STRESS
TEST EQUIPMENT
USED?
CHECK RS INTERFACE
SETTING DETAILED IN
CHAPTER 5
REPLACE: THE MAIN BOARD
CHECK THAT THE CABLE ASSEMBLY TO THE
STRESS TEST EQUIPMENT IS CORRECTLY
AND FIRMLY CONNECTED TO THE RS-232
INTERFACE
ONLY DIGITALLY
CONTROLLED
ERGOMETERS CAN BE
USED WITH THE AT-102