5.7
Article Number 2. 540 028
SCHILLER AG 2002
AT-102 ECG Service Handbook
Section 5
Adjustments
Battery Charge Voltage
Precautions and Requirements
The unit must be placed on an antistatic mat and antistatic precautions observed when any
maintenance is carried out on the AT-102. The room temperature should be between 18 and
28 degrees.
Tools and Equipment
•
Digital voltmeter
•
Small flat bladed screwdriver
•
Resistor 2.7kOhms, 250mW
Procedure
The battery charge voltage is nominally 13.5V. Adjust and check as follows:
1.
Disassemble the unit as detailed in chapter 4.
2.
Remove the Main board MK 18 - 1 from the assembly and remove the EMC shield
from the board.
3
Replace the board in position.
4.
Remove the two battery connectors (if not already removed) and connect a 2.7 kOhm
250mW resistor across the two battery connectors on the main board to simulate a
discharged battery. Connect the digital multimeter across the resistor.
5.
Connect the Mains supply.
WARNING
Exercise care - be aware that potentially lethal voltages are present
6.
Adjust VR3 to obtain a charging voltage of 13.5V.
7.
Disconnect the mains supply and reassemble the unit as detailed in Chapter 4.
Summary of Contents for AT-102
Page 18: ...xviii SCHILLER AT 102 ECG Recorder Service Handbook ...
Page 26: ...1 8 AT 102 Service Handbook Introduction Keypad 4 7 8 9 10 11 12 13 14 15 16 1 2 3 5 6 17 ...
Page 54: ...1 36 AT 102 Service Handbook ...
Page 62: ...2 8 AT 102 Service Handbook ...
Page 78: ...3 16 AT 102 Service Handbook ...
Page 98: ...5 2 AT 102 Service Handbook ...
Page 114: ...5 18 AT 102 Service Handbook ...
Page 118: ...6 4 AT 102 Service Handbook ...
Page 120: ...7 2 AT 102 Service Handbook ...
Page 126: ...7 8 AT 102 Service Handbook ...
Page 134: ...Annex A 8 AT 102 Service Handbook ...
Page 136: ...Index 1 Article Number 2 540 028 SCHILLER AG 2002 AT 102 Service Handbook Index Index ...
Page 140: ...Index 5 Article Number 2 540 028 SCHILLER AG 2002 AT 102 Service Handbook Index ...