3.6
AT-102 Service Handbook
AT-102 Fault Diagnosis Chart (Sheet B - Power Problems)
NO
YES
REMOVE THE TOP
ASSEMBLY. VISUALLY
INSPECT ALL CABLE
ASSEMBLIES AND
CONNECTORS. ENSURE
THAT ALL CONNECTORS
ARE SECURE. CHECK THE
CABLE ASSEMBLY FROM
THE POWER SUPPLY TO
THE TOP ASSEMBLY.
ARE ALL THE CABLE
ASSEMBLIES AND
CONNECTORS OK?
NO
YES
CONNECT A DVM
ACROSS THE MAINS
TRANSFORMER.
REPLACE THE MAINS
TRANSFORMER
NO
YES
ARE ALL POWER
SUPPLIES OK?
REPLACE THE
MAIN BOARD
RECTIFY THE CAUSE OF
THE PROBLEM. REPLACE
FAULTY COMPONENT(S).
IS THE AC VOLTAGE
READING APPROX.
18 V?
POSSIBLE FAULT IN THE
POWER SUPPLY. CHECK ALL
POWER RAILS.
1. REPLACE THE CABLE
ASSEMBLY BETWEEN THE
MAIN BOARD AND THE LCD
(BASE ASSEMBLY TO TOP
ASSEMBLY).
2. REPLACE THE MAIN
BOARD
3. REPLACE THE LCD.
Summary of Contents for AT-102
Page 18: ...xviii SCHILLER AT 102 ECG Recorder Service Handbook ...
Page 26: ...1 8 AT 102 Service Handbook Introduction Keypad 4 7 8 9 10 11 12 13 14 15 16 1 2 3 5 6 17 ...
Page 54: ...1 36 AT 102 Service Handbook ...
Page 62: ...2 8 AT 102 Service Handbook ...
Page 78: ...3 16 AT 102 Service Handbook ...
Page 98: ...5 2 AT 102 Service Handbook ...
Page 114: ...5 18 AT 102 Service Handbook ...
Page 118: ...6 4 AT 102 Service Handbook ...
Page 120: ...7 2 AT 102 Service Handbook ...
Page 126: ...7 8 AT 102 Service Handbook ...
Page 134: ...Annex A 8 AT 102 Service Handbook ...
Page 136: ...Index 1 Article Number 2 540 028 SCHILLER AG 2002 AT 102 Service Handbook Index Index ...
Page 140: ...Index 5 Article Number 2 540 028 SCHILLER AG 2002 AT 102 Service Handbook Index ...