3.7
Article Number 2. 540 028
SCHILLER AG 2002
AT-102 Service Handbook
Section 3
Fault Finding
YES
AT-102 Fault Diagnosis Chart (Sheet C - General Problems)
NO
ARE ALL THE CABLE
ASSEMBLIES AND
CONNECTORS OK?
RECTIFY THE CAUSE OF
THE PROBLEM. REPLACE
FAULTY COMPONENT(S).
NO
YES
ARE ALL POWER
SUPPLIES OK?
REPLACE THEMAIN BOARD
CHECK ALL POWER
RAILS
REMOVE THE TOP ASSEMBLY. VISUALLY
INSPECT ALL CABLE ASSEMBLIES AND
CONNECTORS. ENSURE THAT ALL
CONNECTORS ARE SECURE.
CHECK THE LCD CONTRAST CONTROL.
CHECK THE CABLE ASSEMBLY BETWEEN
THE POWER SUPPLY AND THE LCD
(BOTTOM ASSEMBLY TO TOP ASSEMBLY).
REPLACE REPLACE THE LCD. REPLACE THE
VIDEO/LCD CONVERTER (MAIN BOARD)