5.17
Article Number 2. 540 028
SCHILLER AG 2002
AT-102 ECG Service Handbook
Section 5
Adjustments
Upgrading the Unit / Updating the
Software
5.
Install the SWUP program on the PC (available from the SCHILLER Mailbox).
6.
Follow the instructions given by the program.
7.
Open the SWUP program through the SWUP icon on the desktop or via the Startup -
Schiller-Menu. The following is displayed:
8.
Click the config icon. The following screen is displayed:
9.
Select the applicable COM port (1 or 2). Set the Baudrate to `Auto` (or to the same
Baud rate as set in the AT-102.
10.
Click on `update version` to update the system software.
11.
A series of beeps indicates that the software has been successfully updated. The
updated software can be used immediately.
To exit the
program
without
updating
click the
close icon
Summary of Contents for AT-102
Page 18: ...xviii SCHILLER AT 102 ECG Recorder Service Handbook ...
Page 26: ...1 8 AT 102 Service Handbook Introduction Keypad 4 7 8 9 10 11 12 13 14 15 16 1 2 3 5 6 17 ...
Page 54: ...1 36 AT 102 Service Handbook ...
Page 62: ...2 8 AT 102 Service Handbook ...
Page 78: ...3 16 AT 102 Service Handbook ...
Page 98: ...5 2 AT 102 Service Handbook ...
Page 114: ...5 18 AT 102 Service Handbook ...
Page 118: ...6 4 AT 102 Service Handbook ...
Page 120: ...7 2 AT 102 Service Handbook ...
Page 126: ...7 8 AT 102 Service Handbook ...
Page 134: ...Annex A 8 AT 102 Service Handbook ...
Page 136: ...Index 1 Article Number 2 540 028 SCHILLER AG 2002 AT 102 Service Handbook Index Index ...
Page 140: ...Index 5 Article Number 2 540 028 SCHILLER AG 2002 AT 102 Service Handbook Index ...