PR5230 Instrument Manual
Installing the Instrument and Plug-in Cards
Sartorius
EN-41
Mounting the Cable
Caution!
The connection of the cable screen of the fieldbus cards:
essential see according Chapter 4.3.4 …4.3.9!
1 2 3
4
5 6
Caution!
The cable screen (4) must be connected in the metal sleeve (6) of the cable gland.
Don’t take the screen (4) into the unit!
Before, during and after installation, make sure that the sealing ring is seated
correctly.
Remove the (screw) cap (1).
Slide cap (1) and plastic cone (3) over the cable (2).
Take the cable (5) through the metal sleeve (6).
Fold the cable screen (4) over the lower part of the plastic cone (3) (approx. 10 mm).
Connect the cable conductors.
Tighten the (screw) cap (1).
Tighten metal sleeve (6) inclusive o-ring by lock nut (inside the housing).
Caution!
Regularly check the cable gland for tightness and re-tighten it, if necessary.