Commissioning
PR5230 Instrument Manual
EN-100
Sartorius
5.4.5
Determining the Dead Load
WP A/
Calibration
Max
3000 d
3000 kg
Scale interval
3000 d
1 kg
Deadload at
0.00 kg
0.000000
Max at
3000.00 kg
1.000000
Not calibrated
Sensitivity
833.33
4.000000
by load
by mV/V
CalcTest
To use the empty scale/hopper as dead load (normal case):
discharge the scale/hopper.
press [by load].
After
or confirmation of the change is displayed:
Set deadload
If the mV/V value of the dead load was calculated, or if it is known from the previous calibration, the value can
be overwritten by pressing [by mV/V].
WP A/
Calibration
Max
3000 d
3000 kg
Scale interval
3000 d
1 kg
Deadload at
0.00 kg
0.000000
Max at
3000.00 kg
1.000000
Calibrated at
3000.00 kg
1.000000
Sensitivity
833.33
4.000000
by load
by mV/V
CalcTest