Commissioning
PR5230 Instrument Manual
EN-108
Sartorius
WP A/
Calibration
Standstill range
1.00 d
Tare timeout
2.5 s
Zeroset range
50.00 d
Zerotrack range
0.25 d
Zerotrack step
0.25 d
Zerotrack time
0.0 s
[Standstill range]
The mechanical stability of the scale can be detected as long as any changes in the weight value are within
this range; permissible range 0.01…10.00d. In ‘legal-for-trade’ mode, select
1 d.
[Tare timeout]
Enter a timeout value between 0.1 and 25 s for a taring/zero set command that cannot be executed (e.g., if
scale mechanically unstable, filter settings faulty, resolution too high, stability condition too narrow).
[Zeroset range]
Determine a ±range around the zero point determined by the dead load during calibration; within this range
-
the displayed gross weight can be set to zero by pressing the zero-setting key (or by a corresponding
external command), and
-
automatic zero tracking is active.
Available range: 0.00…10000.00d
In 'legal-for-trade' mode a value
2 % of max. must be entered. Example: 60 d for 3000e, class III.
[Zerotrack range]
Range within which automatic zero tracking compensates deviations, 0.25…10000.00d.
In 'legal-for-trade' mode a value of 0.25 d has to be entered.
[Zerotrack step]
If a weight change exceeds the adjusted value, automatic tracking does not function any more. Step for
automatic tracking 0.25…10d. In 'legal-for-trade' mode a value of 0.25 d has to be entered.
[Zerotrack time]
Enter a time interval for automatic zero tracking within. At 0.0 s the tracking is switched off.
In 'legal-for-trade' mode a value of 1 s has to be entered.
WP A/
Calibration
Overload
9 d
Min 50
d
Range mode
Single range
[Overload]
Weighing range above the maximum capacity [Max], without error message. Available range: 0…999900 d.
In 'legal-for-trade' mode max. 9 d = e has to be entered.
[Min]
Weighing range above the maximum capacity (Max), without error message. Available range: 0…999900 d.
In 'legal-for-trade' mode min. 20 d has to be entered.
[Range mode]
For scale range selection, see Chapter 5.4.12.2.