PR5230 Instrument Manual
Commissioning
Sartorius
EN-117
5.6.4
Printing parameter
Define the basic printing parameters under
-[
Printing parameter
].
Press
to select the specific parameter.
Press
to select the respective values.
Setup/
Printing parameter
Print mode
Cyclic
Cyclic, Cyclic with enable, Triggered
Printing interval
5
Enter value
Printing interval unit
Seconds
Seconds, Minutes, Hours, Measures
PrintlayoutItem1 Sequencenumber
PrintlayoutItem2 Grossweight
PrintlayoutItem3 Date
&
Time
PrintlayoutItem4 CR/LF
PrintlayoutItem5 Form
feed
PrintlayoutItem6 -none-
[Cyclic]
Cyclic printing with adjusted parameters.
[Cyclic with enable]
Continuous printing.
[Triggered]
Singular printing.
[Sequencenumber]
Current sequence number, max. 6 digits, after #999999 the #000001 is following.
[CR/LF], [Form feed]
Carriage returns and line feed, form feed.
[Device address]
Address of the instrument (A, B ... Z).
[-none-]
Nothing is printed. The function can be used for printing less than 6 data items.
[Displayed weight], [Grossweight], [Net weight], [Tare weight]
The displayed gross, net or tare weight is printed. If [OIML], [NTEP] or [NSC] has been selected, printing is done
only, if the stability criteria is fulfilled, the weight is shown in '< >'. For [NTEP] or [NSC] the gross weight is
indicated with G (else B).